Hydrocarbon Analyzer Operating Principle
Other Packing Machine,Vacuum Skin Packaging Machine,Vacuum Packers,Vacuum Machine For Packing WENZHOU BROTHER MACHINERY CO., LTD. , https://www.chinabrother.net
The hydrocarbon analyzer uses a controlled current coulometric analysis method. The water generated by the reaction is carried into the Pt-P2O5 electrolytic cell by the carrier gas (oxygen) and reacts with P2O5 to form metaphosphoric acid. The partial electrolysis of the phosphoric acid is terminated when the electrolytic current is reduced to the end current. electrolysis. A single-chip microcomputer integrates the electricity consumed in the electrolysis process and converts the integrated value of the electricity in real time into the mass of hydrogen and carbon (milligrams). The percentage of carbon and hydrogen is finally displayed and the measurement result is calculated as air. Dry base and dry base forms are printed. When there is no air drying base moisture data, the total hydrogen value and the air dry base carbon value are printed. Related Information: Hydrocarbon Analyzer, Fluorescent Sulfur Detector, Luminescent Nitrogen Detector, Fluorescent Sulphur Meter.